Finding
Paper
Abstract
The impact of 3D rotational errors in patient positioning on dose delivered target volumes and critical structures in IMRT was studied. Patient rotational errors ranging from −30 to +30 was introduced to IMRT treatment plans of pelvis, head and neck and brain treatment sites and the impact of rotational error on DVH metrics was assessed. The magnitude of impact of rotational error on the error in dose delivered to the target volume and critical structures depends on the location of the structures from plan isocentre. In studied plans, a maximum percentage difference of up to −9.8(1s=13.4) % in D95 to PTV was observed for head and neck treatments. Similarly, in Brain treatments a maximum difference of up to 24.0(1s=33.0) % in maximum dose of Optic chiasm was observed. The results suggest that failure to correct patient's rotational error results in under-dosage to target volumes and over-dosage to the critical structures in some specific treatment scenarios.
Authors
S. Arumugam, A. Xing, P. Vial
Journal
Journal of Physics: Conference Series