The mechanical properties of erythrocyte membrane have been studied using an atomic force microscope. Measurements were carried out on blood samples taken from 7 diabetes mellitus patients and 8 healthy individuals. For each blood sample a distribution of a Young's modulus was constructed. It has been found that both the mean value and the width of the distribution in diabetic patients exceed the corresponding results for healthy persons by a factor greater than 3. The high sensitivity of the atomic force microscopy and the ability to measure the full distribution of the erythrocyte membrane Young's modulus makes it a unique, powerful and promising tool in studies of the membrane stiffness of red blood cells.