Technical note: radiolucent operating room table extension to facilitate intraoperative angiography.
Published 1991 · Stephen T. Hecht, Susan S. Kemp, C. W. Kerber
AJNR. American journal of neuroradiology
5
Citations
1
Influential Citations
Abstract
Recent improvements in digital C-arm radiographic units permit the safe and effective performance of cerebral angiography in the operating room. Intraoperative cerebral angiography, in turn, facilitates more effective aneurysm surgery and arteriovenous malformation therapy (1 , 2]. A variety of radiolucent carbon fiber head holders are now available for use in neurosurgical procedures that involve angiography. Radiolucent operating room tables, suitable for angiography, are also now commercially available. Most operating room tables currently in active use, however, have been designed without consideration of possible use as angiography tables. Radiographic visualization of the aortic arch is essential for the performance of cerebral angiography, and radiopaque tables obscure this structure. During angiography on radiopaque tables, the aortic arch can often be visualized by proper angulation of the C-arm unit, although the resulting projections are disorienting. Frontal viewing of the aortic arch is made possible with any operating room table by the addition of a durable, weightbearing, radiolucent table extension that can be easily and inexpensively fabricated. The patient's chest is supported by the radiolucent extension , which projects over the edge of the radiopaque table. At present, there are no commercially available radiolucent table extensions to adapt radiopaque operating room tables for cerebral angiography. The extension can be made from ~-in . carbon fiber sheeting, Sfa-in . Plexiglas, or 3ftt-in . plywood. A 2 x 5 ft sheet of the radiolucent material (carbon fiber, Plexiglas, or plywood) is placed on the operating room table with approximately 2025 em projecting off the head end (Fig. 1 ). A soft pad is placed over the sheet of radiolucent material, which is then held in place by the weight of the patient. Prior to attaching the head holder, test the system by pressing on the head end of the board to be certain it is stable. To avoid unintended motion, the extension board is taped to the operating room table. The Fig. 1.-Three-quarterin. plywood board on operating room table. Board is covered by foam rubber pad. Head holder is attached to operating room table. Note that elbow joint of head holder (arrow) is in an extended position, allowing increased clearance for C-arm.
A radiolucent table extension can be easily and inexpensively fabricated to enable frontal viewing of the aortic arch during intraoperative cerebral angiography on radiopaque operating room tables.
Full text analysis coming soon...