J. Komarek, R. Červenka, T. Růžička
Nov 5, 2007
Citations
2
Influential Citations
42
Citations
Journal
Journal of pharmaceutical and biomedical analysis
Abstract
Conditions of a continuous flow extraction (CFE) of aluminium acetylacetonate in acetylacetone and aluminium 8-hydroxyquinolinate into methylisobutylketone (lengths of reaction and extraction coils, flow rates of aqueous and organic phases and their flow rate ratio, pH of aqueous phase, lengths of coils for transport of aqueous and organic phases and effect of salts) were studied. The analytical signal of the aluminium chelates present in the organic phase was measured at 309.3 nm using atomic absorption spectrometry with electrothermal atomization (ET-AAS) at the flow rate ratio F aq/F org=3 for aqueous and organic phases. The five points calibration curves were linear (R2 0.9973 and 0.9987) up to 21 microgl(-1) Al with the limits of detection of 0.3 microgl(-1) and the recovery 100+/-2% and precision of 3% at 2-10-fold dilution of the dialysis concentrates. The acetylacetonate method was applied to the determination of aluminium in real dialysis concentrates. Aluminium in concentrations 5-6 microgl(-1) (R.S.D.s 5-10% in real samples) were found and the results were in the very good agreement with those obtained by an ET-AAS using preconcentration of Al(III) on a Spheron-Salicyl chelating sorbent (absolute and relative differences were under 0.4 microgl(-1) and 8.2%, respectively).