S. Mittal, S. A. Kumar, H. Sharma
Mar 10, 2004
Citations
0
Influential Citations
82
Citations
Quality indicators
Journal
Talanta
Abstract
A new ion-selective electrode (ISE) based on dicyclohexano-18-crown-6 (DC18C6) as a neutral carrier is developed for lanthanum(III) ions. The electrode comprises of dicyclohexano-18-crown-6 (6%), PVC (33%), and ortho-nitrophenyl octyl ether (o-NPOE) (61%). The electrode shows a linear dynamic response in the concentration range of 10(-6) to 10(-1)M with a Nernstian slope of 19mV per decade and a detection limit as 5x10(-7)M. It has a response time of <30s and can be used for at least 5 months without any significant divergence in potentials. The selectivity coefficients for mono-, di-, and trivalent cations indicate good selectivity for La(III) ions over a large number of interfering cations. The sensor has been used as an indicator electrode in the potentiometric titrations of La(III) with EDTA. The membrane is successfully applied in partially non-aqueous medium. It can be used in the pH range 4-9.