Theano D. Karakosta, P. Tzanavaras, D. G. Themelis
Aug 1, 2011
Citations
0
Influential Citations
2
Citations
Journal
Journal of separation science
Abstract
The present study reports the very first application of ethyl propiolate (EP) as an advantageous pre-column derivatization reagent for the determination of thiols by liquid chromatography (LC). Cysteine (CYS), glutathione (GSH) and N-acetylcysteine (NAC) were derivatized online under stopped-flow conditions in a sequential injection (SI) system coupled to HPLC. The formed derivatives were separated isocratically with a monolithic stationary phase (100×4.6 mm id) and UV detected at 285 nm. Critical parameters that affected the online pre-column derivatization reaction (e.g. the reaction time and the amount concentration of EP) and the separation (e.g. pH and the composition of the mobile phase) were investigated. The developed analytical scheme offers a total analysis time of less than 10 min, limits of detection in the range of 0.24-0.35 μmol/L and satisfactory linearity up to 200 μmol/L for all analytes. The proposed method was applied to the analysis of the selected thiols--that are often employed as antibrowning agents--in fresh fruit samples.