Xiao Chen, Hongling Zhang, Yanjian Wan
Jan 17, 2018
Citations
0
Influential Citations
16
Citations
Quality indicators
Journal
PLoS ONE
Abstract
2,4-Dichlorophenoxyacetic acid (2,4-D) is a chlorophenoxy herbicide used worldwide. We describe a high-performance liquid chromatography (HPLC) method with UV detection for the determination of 2,4-D in female and male rat serum. This allows to observe the change of serum 2,4-D concentration in rats with time and its pharmacokinetics characteristics with a simple, rapid, optimized and validated method. The serum samples are pretreated and introduced into the HPLC system. The analytes are separated in a XDB-C18 column with a mobile phase of acetonitrile (solvent A) and 0.02 M ammonium acetate (containing 0.1% formic acid) (solvent B) using a gradient elution at a flow rate of 1.0 mL/min. The wavelength for UV detection was set at 230 nm. Calibration curve for 2,4-D was constructed over a range of 0.1–400 mg/L. The method was successfully applied to study the pharmacokinetics of 2,4-D in rats in this study. After oral administration of 300 mg/kg and 60 mg/kg 2,4-D, the mean Cmax values were 601.9 and 218.4 mg/L, the AUC0→∞ values were 23,722 and 4,127 mg×h/L and the clearance (Cl) were 1.10 and 0.02 L/(h×kg), respectively. The developed method was found to be specific, precise, reproducible and rapid.